Jonathan Pearce
National Physical Laboratory
UK
Thermometers have widespread application and like any measurement device, they must be calibrated. Many do this with the International Temperature Scale of 1990 (ITS-90), which is a platinum resistance thermometer that calibrates using a fixed point cell, which is a container with a well running through its center for insertion of the thermometer. The container has a material of 99.9999% purity. The central well allows a thermometer to be inserted so that the sensing element at its bottom is completely immersed in the fixed point metal. The container is then placed in a furnace to allow controlled melting and solidification of the metal.
To better understand the microscopic behavior of the phase change processes, researchers at the National Physical Laboratory turned to simulation. They were able to identify specific behavior in a system that is difficult to observe experimentally and yet contributes to the overall uncertainty in a measurement. This showed that the liquid-solid interface is not planar and ripples become apparent that become the cells and begin to protrude outward. The tips of these cells have a significantly lower temperature than their roots. The information gained through simulation can be applied to a thoughtful re-design of the device, ultimately improving measurement precision.
Suggested Stories
数値シミュレーションでデンドライトを制御する
効率的に機能するリチウムイオン電池を設計するために Murata Manufacturing Co., Ltd. のエンジニアは電池内のリチウム金属電極に電流を加えたときに発生する樹状突起の形成を評価し, ...
マルチフィジックスシミュレーションによる高電圧分布のためのGISの最適化
ガス絶縁型金属封入スイッチギア(GIS)の開発プロセスをスピードアップする一方でコストを大幅に削減するため, Pinggao Group はCOMSOLマルチフィジックス®を使用しました. ...
Get a Quick Look at How Plastometrex Uses COMSOL Compiler™
Step into the world of the startup company Plastometrex to see how they developed an efficient method for measuring ...
リードレス心臓ペースメーカーの体外通信の検討
MicroPort CRMは, 数値シミュレーションを使用して, リードレス心臓ペースメーカー(LCP)の設計と, 患者の心拍管理のために人体内で相互に通信する方法を最適化しました.